Hardness(Ca,Mg) Analyzer
-
1Convenient maintenance
- Easy tube replacement & disassembly
- Simplified structure
- Data storage function (USB)
-
27” Color LCD Monitor
-
3Display of data trend
-
4Grab sample analysis
-
5Automatic calibration by standard solution
-
6Multichannel operation (1∼3ch)
-
7Applications
- For feed water and boiler water quality management of industrial boilers
Technical Specifications
Model | AH-010R |
AH-010N
|
AH-300N | |
---|---|---|---|---|
Measuring Principle | Colorimetric Method | |||
Measuring Range | 0~3.0 mg/L (ppm) | 0~10.0 mg/L (ppm) | 30~300 mg/L (ppm) | |
Display | 0.01 mg/L (ppm) | 0.1 mg/L (ppm) | 1 mg/L (ppm) | |
Accuracy | ±0.05 mg/L |
±0.2 mg/L or±5% |
±5.0 mg/L or ±5% | |
Response Time | 8~15 min | 5~15 min | 5~15 min | |
Calibration | Automatic by standard solution | |||
Sample Pressure | 1~5 bar | |||
Sample Temperature | 5~45 ℃ | |||
Sample Flow | 200~500 ml/min | |||
Reagent Consumption | 60 days (Reagent A, C / 2 liter, Reagent B / 1 liter, Standard solution A, B / 2 liter) | |||
Data Output | 4∼20 mA 2 ea, Alarm 2 ea, RS 485 Modbus (Option) | |||
Panel Dimension (mm) | 400W X 500H X 200D (Flush, Wall mounting), 700W X 1550H X 350D (Standing type) | |||
Degree of Protection | IP65( NEMA 4x) : Electronics, IP55(NEMA 4) : Wet part | |||
Power | 100∼240 VAC (50/60 Hz) |
Analysis Flow
Installation
Dimensions of panel mounting
Dimensions of wall mounting
Consumable List
Part No. | Description | Replacement Cycle | Remarks |
---|---|---|---|
HN01-010 | Reagent pump tube | 3 months | Tube set (A, B) |
HN01-011 | Fitting tube | 6 months | Reagent injection tube |
HN01-012 | Reagent bottle fitting | 1 year | Fitting and tube in bottle |
HN01-016 | Reagent A | 2 months | Reagent (2 liter) |
HN01-017 | Reagent B | 2 months | Reagent (1 liter) |
HN01-018 | Reagent C | 2 months | Reagent (2 liter) |
HN01-019 | Standard solution A | 2 months | High Point (2 liter) |
HN01-020 | Standard solution B | 2 months | Low Point (2 liter) |
HN01-022 | Pump (Standard solution) | Failure | Supply pump for standard solution or grab sample |
HN01-023 | Pump (Reagent) | Failure | Supply pump for reagent |
HN01-035 | Sample filter | For preventing impurities inflow |