Sulfite(SO3) Analyzer

-
1Convenient maintenance
- One-touch automatic operation of various menus
- Easy tube replacement & disassembly
- Data storage function (USB)
-
27” Color LCD Monitor & Touch Screen
-
3Display of data trend
-
4Automatic calibration by standard solution
-
5Automatic cleaning function
-
6Applications
- Process of exhaust gas desulfurization
facilities in power plants
- Sulfite
concentration monitoring to optimize biocide injection
and prevent pipeline corrosion in geothermal power plant cooling systems
- Process of exhaust gas desulfurization
facilities in power plants
Technical Specificatins
| Model | ASP-250 |
|---|---|
| Measuring Principle | Colorimetric Method |
| Measuring Range | 0~30 mg/L(ppm) / 100~12,000 mg/L(ppm) |
| Display | 0.1 mg/L(ppm) / 1 mg/L(ppm) |
| Accuracy | ±0.5 mg/L or ±5 % / ±10 mg/L or ±2 % |
| Response Time | 15
min App. (Cycle time : 30 min) |
| Calibration | Automatic
by standard solution |
| Sample Pressure | 1∼5 bar |
| Sample Temperature | 5∼45℃ |
| Sample Flow | 200∼700
ml/min |
| Sample Connection | Inlet 6mm/ Outlet 16mm |
| Reagent Consumption | 30
days (Reagent A / 1
liter, Reagent B / 1 liter, Standard Solution /
2 liter) |
| Data Output | 4∼20 mA 2 ea, Alarm 2 ea |
| Panel Dimension (mm) | 400W X 500H X 200D (Flush, Wall mounting), 700W X 1620H X 350D (Standing type) |
| Degree of Protection | IP65(
NEMA 4x) : Electronics, IP55(NEMA 4) : Wet part |
| Power | 100∼240VAC(50/60Hz) |
Analysis Flow
Installation
Standing type
Consumables and Spare parts list
| Part No. | Description | Replacement Cycle |
Remarks |
|---|---|---|---|
| SP01-010 | Reagent pump tube | 2 months | Tube set (A, B) |
| SP01-011 | Teflon coating tube (2M) | 6 months | For reagent injection |
| SP01-012 | Reagent bottle cap set | 1 year | |
| SP01-014 | Champber injection sample tube (1m) | 6 months | |
| SP01-016 | Reagent A | 30 days | Reagent (1 liter) |
| SP01-017 | Reagent B | 30 days | Reagent (1 liter) |
| SP01-019 | Standard material (50g) | 30 days | 10 ppm, Prepared by SO₃ standard solution method provided by KNTEC |
| SP01-022 | Pump (Standard solution) | 1 year |
|
| SP01-023 | Pump (Reagent) | Failure | |
| SP01-024 | Electromagnetic
flow meter |
Failure |
Monitor for sample flow rate |
| SP01-025 | 3 Way solenoid valve | Failure | For multi-channel |
| SP01-032 | Sample injection pump tube | 6 months | |
| SP01-043 | Sample flow meter (700cc) | Failure |